FDA Recall Open, Classified

7.5MHZ Endo Transducer Probe

Recall: Z-2348-2025 · Initiated July 3, 2025

Recall

Recall Number
Z-2348-2025
Event Number
97217
Firm
Philips Ultrasound, Inc
FEI Number
2518586
Product Code
ITX
Status
Open, Classified
Root Cause
Device Design
Initiated
July 3, 2025
Posted
August 22, 2025
Address
1 Echo Dr, Reedsville, PA, 17084-8603

Description

7.5MHZ Endo Transducer Probe

Reason

To provide clarification and labelling to define the useful life of ultrasound transducers in the field.

Action

An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.

Distribution

US Nationwide distribution.

Quantity

171,322 units