FDA Recall Terminated

Ingenia Elition S-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781357 (OUS)

Recall: Z-2347-2021 · Initiated July 30, 2021

Recall

Recall Number
Z-2347-2021
Event Number
88461
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
LNH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 30, 2021
Terminated
July 25, 2023
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Ingenia Elition S-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781357 (OUS)

Reason

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Action

Philips North America issued U.S. customers letters via certified mail using a 3rd Party service. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall , health risk and action to take: The Philips Ingenia 1.5T, Ingenia 1.5T Evolution, Ingenia 3.0T, and Ingenia Elition X MR systems can continue to fulfill clinical use as normal. However, in case of technical problems related to off cooling of the magnet, magnet pressure or Helium boil-off, please stop use and contact your local Philips service representative immediately and reference FCO78100530. A Philips Field Service Engineer (FSE) will contact you to schedule an on-site visit to resolve the issue by inspecting the potentially affected system(s) to confirm if the non-conforming pressure relief (42 psi) has been installed (reference FCO78100530). If your system has a 42 psi valve, the Phillips FSE will replace the valve with the correct 3.75 psi valve. you need any further information or support concerning this issue, please contact the Customer Care Solutions Center 1-800-722-9377 and reference FCO78100530.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.

Quantity

7 units