FDA Recall Terminated

Service parts for Crank Screw Assembly (PN1069-042-010S and 1069-042110S) Stryker Medical 3800 E. Centre Ave Portage MI 49002. Intended use: Service parts are replacement parts for the eye surgery stretcher.

Recall: Z-2343-2010 · Initiated February 26, 2010

Recall

Recall Number
Z-2343-2010
Event Number
54265
Firm
Stryker Medical Div. of Stryker Corporation
FEI Number
1831750
Product Code
FNJ
Status
Terminated
Root Cause
Device Design
Initiated
February 26, 2010
Posted
September 1, 2010
Terminated
April 21, 2011
Address
3800 E. Centre Ave., Portage, MI, 49002

Description

Service parts for Crank Screw Assembly (PN1069-042-010S and 1069-042110S) Stryker Medical 3800 E. Centre Ave Portage MI 49002. Intended use: Service parts are replacement parts for the eye surgery stretcher.

Reason

The fowler of the bed is raised by a crank. The crank consists of a steel screw that moves through a brass nut. Over time the screw can wear the threads on the brass nut. The screw and the nut can become misaligned, causing a loud squealing noise when the screw comes in contact with the drive tube. In addition the fowler can become inoperable over time with excessive wear of the brass nut. T

Action

The firm, Stryker Medical, sent an "Urgent Medical Device Correction" letter dated February 26, 2010, to all customers. The letter describes the product, problem and action to be taken by customers. The letter states that Stryker field service tech will contact the firm and perform device corrections within four months. The customers were instructed to locate the recalled stretchers and verify that there is no grinding noise when activating the fowler. If the fowler exhibited a grinding noise, the customers were instructed to remove the stretcher from service and contact the recalling firm at 800-STRYKER, option 3. Note: Do not put the stretcher back into service until it is repaired by our field service representative. The customers were also instructed to return the enclosed post card to confirm receipt of this notice; inform any new users and advise the firm of new locations if any of the listed stretchers were on loan or sold, and advise the firm of the obsolescence by providing the serial numbers if any of the listed stretchers were disposed of. If you have any urgent questions or concerns, contact (269) 324-6884. Normal hours are Monday-Friday 8am-5pm(EST).

Distribution

Worldwide distribution: USA and countries including Australia, Canada, Chile, China, France, Germany, Greece, Spain, Italy, Japan, South Africa, Taiwan, Korea, Singapore, and UK.

Quantity

226