Eye surgery stretcher with crank fowler, model 1069 Stryker Medical 3800 E. Centre Ave Portage MI 49002. Intended use: A stretcher used for eye surgery.
Recall
- Recall Number
- Z-2341-2010
- Event Number
- 54265
- Firm
- Stryker Medical Div. of Stryker Corporation
- FEI Number
- 1831750
- Product Code
- FNJ
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- February 26, 2010
- Posted
- September 1, 2010
- Terminated
- April 21, 2011
- Address
- 3800 E. Centre Ave., Portage, MI, 49002
Description
Eye surgery stretcher with crank fowler, model 1069 Stryker Medical 3800 E. Centre Ave Portage MI 49002. Intended use: A stretcher used for eye surgery.
The fowler of the bed is raised by a crank. The crank cosnsists of a steel screw that moves through a brass nut. Over time the screw can wear the threads on the brass nut. The screw and the nut can become misaligned, causing a loud squealing noise when the screw comes in contact with the drive tube. In addition the fowler can become inopperable over time with excessive wear of the brass nut.
The firm, Stryker Medical, sent an "Urgent Medical Device Correction" letter dated February 26, 2010, to all customers. The letter describes the product, problem and action to be taken by customers. The letter states that Stryker field service tech will contact the firm and perform device corrections within four months. The customers were instructed to locate the recalled stretchers and verify that there is no grinding noise when activating the fowler. If the fowler exhibited a grinding noise, the customers were instructed to remove the stretcher from service and contact the recalling firm at 800-STRYKER, option 3. Note: Do not put the stretcher back into service until it is repaired by our field service representative. The customers were also instructed to return the enclosed post card to confirm receipt of this notice; inform any new users and advise the firm of new locations if any of the listed stretchers were on loan or sold, and advise the firm of the obsolescence by providing the serial numbers if any of the listed stretchers were disposed of. If you have any urgent questions or concerns, contact (269) 324-6884. Normal hours are Monday-Friday 8am-5pm(EST).
Worldwide distribution: USA and countries including Australia, Canada, Chile, China, France, Germany, Greece, Spain, Italy, Japan, South Africa, Taiwan, Korea, Singapore, and UK.