FDA Recall Terminated

Toshiba Aquilion ONE CT SystemTSX-301 A/2D

Recall: Z-2339-2010 · Initiated February 3, 2009

Recall

Recall Number
Z-2339-2010
Event Number
52830
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
February 3, 2009
Posted
September 1, 2010
Terminated
September 2, 2011
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068

Description

Toshiba Aquilion ONE CT SystemTSX-301 A/2D

Reason

ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB.

Action

Toshiba America Medical System Inc (TAMS) sent an Urgent: Medical Device Correction dated January 14, 2009 to their consignees. The letter described the issue and provided instructions to test the device to determine if it is an affected device. If the device is affected, contact the Toshiba service representative at 800 521-1968. Consignees were requested to complete and fax a response form to Toshiba.

Distribution

Nationwide Distribution: including the states of AL, CA, FL, GA, KY, MA, MD, MT, NV, NY, OH, PA, and Puerto Rico.

Quantity

16