FDA Recall
Terminated
Toshiba Aquilion ONE CT SystemTSX-301 A/2D
Recall: Z-2339-2010
·
Initiated February 3, 2009
Recall
- Recall Number
- Z-2339-2010
- Event Number
- 52830
- Firm
- Toshiba American Medical Systems Inc
- FEI Number
- 2020563
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 3, 2009
- Posted
- September 1, 2010
- Terminated
- September 2, 2011
- Address
- 2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068
Description
Toshiba Aquilion ONE CT SystemTSX-301 A/2D
Reason
ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB.
Action
Toshiba America Medical System Inc (TAMS) sent an Urgent: Medical Device Correction dated January 14, 2009 to their consignees. The letter described the issue and provided instructions to test the device to determine if it is an affected device. If the device is affected, contact the Toshiba service representative at 800 521-1968. Consignees were requested to complete and fax a response form to Toshiba.
Distribution
Nationwide Distribution: including the states of AL, CA, FL, GA, KY, MA, MD, MT, NV, NY, OH, PA, and Puerto Rico.
Quantity
16