Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.

Recall: Z-2338-2025 · Initiated July 25, 2025

Recall

Recall Number
Z-2338-2025
Event Number
97234
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
GEI
Status
Open, Classified
Root Cause
Component change control
Initiated
July 25, 2025
Posted
August 19, 2025
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.

Reason

Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

Action

Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.

Distribution

US Nationwide distribution.

Quantity

155 units