FDA Recall
Terminated
Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating rooms and ER, for patient intubation.
Recall: Z-2338-2008
·
Initiated July 7, 2008
Recall
- Recall Number
- Z-2338-2008
- Event Number
- 48979
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- BSZ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 7, 2008
- Posted
- September 23, 2008
- Terminated
- March 8, 2010
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating rooms and ER, for patient intubation.
Reason
The parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.
Action
The firm has issued a MEDICAL DEVICE RECALL letter dated 7/7/08 to its customers via certified mail. The letter informs them of the problem and to exercise care when using the device until it has been retrofitted. Contact Drager Medical at 1-800-543-5047, #4 for assistance.
Distribution
Nationwide
Quantity
111 units