Ultraview Digital Telemetry Transmitter, Model 91341-09 The Spacelabs Medical Digital Telemetry system provides continuous electrocardiographic (ECG) monitoring to detect ST segment changes, abnormal cardiac rhythms, including asystole, ventricular fibrillation and ventricular tachycardia.
Recall
- Recall Number
- Z-2336-2008
- Event Number
- 48973
- Firm
- Spacelabs Healthcare, Incorporated
- FEI Number
- 3010157426
- Product Code
- DSI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 17, 2008
- Posted
- September 25, 2008
- Terminated
- February 16, 2011
- Address
- 5150 220th Ave Se, Issaquah, WA, 98029-6834
Description
Ultraview Digital Telemetry Transmitter, Model 91341-09 The Spacelabs Medical Digital Telemetry system provides continuous electrocardiographic (ECG) monitoring to detect ST segment changes, abnormal cardiac rhythms, including asystole, ventricular fibrillation and ventricular tachycardia.
During battery insertion there is potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
On 07/17/08 the firm issued a "URGENT - MEDICAL DEVICE CORRECTION" letter to their customers who received the Spacelabs Medical 1400 MHz Telemetry System - Model 91341-09. The letter states the circumstances in which the device malfunctions, instructions for staff to confirm at the time of the battery insertion that the data is appearing and to remove it from service if information is not appearing. Letters will be sent with a return receipt acknowledging each customer receives the Medical Device Correction letter. Spacelabs Technical Support will respond to all technical questions and personally upgrade all switches with resisters. Contact Spacelabs Medical at 1-800-522-7025 for assistance.
Nationwide
2,055 transmitters