FDA Recall Terminated

SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin.

Recall: Z-2334-2012 · Initiated July 8, 2005

Recall

Recall Number
Z-2334-2012
Event Number
61869
Firm
JK Products & Services, Inc
FEI Number
2311923
Product Code
LEJ
Status
Terminated
Root Cause
Equipment maintenance
Initiated
July 8, 2005
Posted
September 7, 2012
Terminated
January 13, 2017
Address
1 Walter Kratz Dr, Jonesboro, AR, 72401-0609

Description

SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin.

Reason

The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor piece, the cord has been known to fail causing a short in the wiring at the pivot point in the cord.

Action

JK Global Services sent out Mandatory Inspection/Modification letters to their Distribution Service Teams dated 07/08/2005, 01/12/2006, 09/15/2010, 10/04/2010 and 03/01/2012. A letter dated September 15, 2010, was sent to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to contact a JK representative or their Distributor as soon as possible to schedule an inspection to the affected product in their possession. For any questions customers were instructed to call 1-800-445-0624 or send an e-mail to [email protected].

Distribution

Worldwide Distribution - USA (nationwide) and Canada

Quantity

1948 units