ZOLL OneStep Multi-Function Electrode; Adult Complete Part #: 8900-0214-01. The electrodes are indicated for the following clinical applications: Defibrillation, Cardioversion, Noninvasive pacing, and Electrocardiograph monitoring.
Recall
- Recall Number
- Z-2334-2008
- Event Number
- 48753
- Firm
- Bio-Detek, Inc.
- FEI Number
- 1218058
- Product Code
- MLN
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- June 16, 2008
- Posted
- September 20, 2008
- Terminated
- May 10, 2016
- Address
- 525 Narragansett Park Dr, Pawtucket, RI, 02861-4323
Description
ZOLL OneStep Multi-Function Electrode; Adult Complete Part #: 8900-0214-01. The electrodes are indicated for the following clinical applications: Defibrillation, Cardioversion, Noninvasive pacing, and Electrocardiograph monitoring.
Multi-function electrode programmed as a Pediatric instead of Adult
Bio-Deteck notified the own label distributor and in cooperation the firm notified their customers by letter dated 6/16/08. Users are requested to return the lots and replacement product will be provided. Zoll issued and Urgent Device Correction letter on June 16, 2008 with instructions to follow if the device is used with the defective electrodes. If you have questions, contact ZOLL Medical Technical Service at 1-800-348-9011.
MA
25 cases of 8/cs for each lot # ; Total of 400 units