FDA Recall Terminated

BD" Mouse IgG2a Isotype Control APC X39 ASR

Recall: Z-2333-2021 · Initiated July 22, 2021

Recall

Recall Number
Z-2333-2021
Event Number
88341
Firm
Becton, Dickinson and Company, BD Biosciences
FEI Number
2916837
Product Code
MVU
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 22, 2021
Terminated
March 24, 2023
Address
2350 Qume Dr, San Jose, CA, 95131-1812

Description

BD" Mouse IgG2a Isotype Control APC X39 ASR

Reason

Complaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.

Action

On 07/22/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification letter via FedEx to customers to inform them that the Recalling Firm received 3 complaints for performance issues related to high background and staining. Our investigation has confirmed there was an error during the manufacturing process that can potentially cause a false positive staining due to contamination with another antibody. Customers are instructed to: 1) Immediately review their inventory for the specific Catalog and lot number listed in the Recall Notification above. Destroy all product subject to the recall following their institution process for destruction. 2) Share this recall notification with all users of the product within their facility to ensure that they are also aware of this recall. If the affected product was used to generate test results, review potential impact to test results following their institution's policies. 3) Complete the attached Customer Response Form and return to the BD contact noted on the form. Even if you do not have any of the affected lots in your inventory, please complete the Customer Recall Response Form indicating you have zero (0) quantity and return as indicated. Actions being taken by the Recalling Firm: 1) An investigation has been initiated to identify root cause and corrective actions to prevent recurrence. 2) The Recalling Firm will provide replacement for any discarded product. For questions or additional assistance contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) M-F 8am -5pm CT

Distribution

Worldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.

Quantity

23 reagents