FDA Recall Completed

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Recall: Z-2332-2024 · Initiated April 24, 2024

Recall

Recall Number
Z-2332-2024
Event Number
94731
Firm
Barco N.V. President Kennedypark 35 Kortrijk Belgium
FEI Number
3004455125
Product Code
DXJ
Status
Completed
Root Cause
Other
Initiated
April 24, 2024
Posted
July 11, 2024

Description

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Reason

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Action

Barco NV notified its sole consignee on 04/29/2024 via email. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by the customer: 1. Identify the location and amount of all Molex modules that are listed in pdf document Molex_SN_Partner . 2. Fill out and send back Response Form 1 to your Barco Key Account Manager, including the number of distributed units per country. 3. Disable or discard all used Molex modules. 4. Fill out and send back Response Form 2 to your Barco Key Account Manager after completion of the actions. For question, contact your Barco Key Account Manager.

Distribution

US: FL

Quantity

88 units