Boston Scientific Ultraflex Precision Colonic Stent System Material/UPN/Catalog Number: M00557360
Recall
- Recall Number
- Z-2331-2010
- Event Number
- 56331
- Firm
- Boston Scientific Corporation
- FEI Number
- 3005099803
- Product Code
- MQR
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 15, 2010
- Posted
- August 31, 2010
- Terminated
- October 14, 2011
- Address
- 100 Boston Scientific Way, Marlborough, MA, 01752-1234
Description
Boston Scientific Ultraflex Precision Colonic Stent System Material/UPN/Catalog Number: M00557360
Units have excessive fraying of the deployment suture thread, as a result, these devices may experience a suture break during deployment.
Boston Scientific issued "Urgent Medical Device Recall - Immediate Action Required" letters dated July 15,2010 to consignees. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. Foreign accounts received notification translated into the local language and with changes that are applicable to local regulations. Boston Scientific can be contacted at 1-508-683-4427.
Worldwide Distribution: USA, and the countries of: Austria, Germany, Finland, France, UK, Ireland, Iran, Italy, the Netherlands, Portugual, Spain, and Sweden
8 units