FDA Recall Open, Classified

AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.

Recall: Z-2330-2021 · Initiated July 9, 2021

Recall

Recall Number
Z-2330-2021
Event Number
88337
Firm
PROCEPT BIOROBOTICS CORPORATION
FEI Number
3027788423
Product Code
PZP
Status
Open, Classified
Root Cause
Component change control
Initiated
July 9, 2021
Address
900 Island Dr, Ste 101, Redwood City, CA, 94065-5176

Description

AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.

Reason

Scope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both tip detachment and entire tip separation form the handpiece may lead to procedural delay for handpiece exchange. Tip detachment can be seen cystoscopically and tip separation from the handpiece may necessitate tip removal from the bladder or prostate.

Action

On 07/09/21 Voluntary Medical Device Recall notices were mailed to customers. Customers were asked to take the following actions if tip detachment occurs: If the scope tube tip detaches from the telescoping tube before insertion into the patient during preparation, the Handpiece must be replaced with a new Handpiece. If the scope tube tip detaches after the Handpiece has been inserted into the patient, this is seen cystoscopically and the Handpiece must be removed and replaced with a new Handpiece. In the event that the scope tube tip separates from the Handpiece entirely, the scope tube tip may fall into the bladder or the prostate. The standard 26Fr resectoscope, used in an Aquablation procedure, will accommodate the removal of the detached scope tube tip. In addition, customers were asked to take the following actions: 1) Read the notice. 2) Stop using affected devices. 3) Remove affected lots from their regular inventory location and segregate for future return and replacement. 4) Complete and return the acknowledgement form. 5) Pass along the notice to all that need to be aware. Customers with questions were encouraged to contact Customer Service: 1.650.232.7222 or e-mail [email protected].

Distribution

Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, WI, AZ, AR, MI, WV, SD, PA, CA, DC, NJ, NY, GA, IN, VA, TN and the countries of Austria, Switzerland, Germany, Spain, France, UK, Italy.

Quantity

608