FDA Recall Open, Classified

Mazor X Surgical System Positioner Type II, REF: ASM0214-02

Recall: Z-2330-2020 · Initiated December 5, 2019

Recall

Recall Number
Z-2330-2020
Event Number
85387
Firm
Medtronic Navigation, Inc.
FEI Number
1000517638
Product Code
OLO
Status
Open, Classified
Root Cause
Device Design
Initiated
December 5, 2019
Posted
June 9, 2020
Address
826 Coal Creek Cir, Louisville, CO, 80027-9710

Description

Mazor X Surgical System Positioner Type II, REF: ASM0214-02

Reason

There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]

Action

The firm originally sent a "Medical Device Correction " letter to affected consignees on December 5, 2019. On November 9, 2020, a "Update - Medical Device Correction" letter was sent out to all affected customers. The purpose of the letter was to update consignees that since the initiation of the recall (12/5/2019), Medtronic had developed a fix to permanently correct the issue associated with the Mazor X Surgical system. The notice further stated that upon availability of parts, Medtronic will schedule a service visit to correct the system at no charge. Medtronic expects to initiate these service visits within the next six months. Pending completion of service to correct the Positioner Type II system, continue to follow the modified mounting instructions. The firm requested that consignees do the following: 1. Please sign and date the Customer Confirmation Form enclosed with this letter to confirm that you have reviewed the information included in this notification with all users of the affected Mazor X" Surgical System, including physician users. 2. Return the form to Medtronic via email at [email protected] or via fax at 651-367-7075 within 30 days of receipt. 3. If you have any questions regarding this notification, please contact Medtronic at 1-888-826-5603. Medtronic will initiate service visits to correct systems on-site at customer locations when the kits become available. This is expected to begin by February 2021. View spreadsheet for updated

Distribution

US: ALAR,AZ,CA,CO,CT,FL,GA,IA,ID,IL,IN,KY,MA,MI,MN,MO,NC,ND,NL,NY,OH,PA,SC,TN,TX,UT,VA,WA,WI,WV, OUS: Australia, China, India, Israel, Romania, Singapore, Thailand, United Kingdom, Switzerland, Spain, Germany, Belgium,

Quantity

177 Systems