FDA Recall Terminated

EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.

Recall: Z-2330-2016 · Initiated July 6, 2016

Recall

Recall Number
Z-2330-2016
Event Number
74608
Firm
Dako Denmark A/S Produktionsvej 42 Glostrup Denmark
FEI Number
3002806778
Product Code
LDT
Status
Terminated
Root Cause
Process control
Initiated
July 6, 2016
Terminated
October 13, 2016

Description

EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.

Reason

The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.

Action

A recall notification letter dated 7/6/16 was sent to customers to inform them that Dako is recalling EnVision FLEX+ kit because the pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1. The letter informs the customers that with a pH of 4-5 instead of pH 9 there could be weak staining, false negative or false positive non-specific staining depending on the antibody used. There is the possibility that this staining could lead to a false result. Customers are to discard the affected product, determine the impact as instructed, and confirmed that they have received replacement kits. Customers are instructed to complete the Device Recall Form and contact local sales representative for any questions regarding the recall notification.

Distribution

Worldwide Distribution -- US, Canada, and Colombia.

Quantity

147 kits