FDA Recall Terminated

S2 Fluid Pack on the Cobas b 221 system- Catalog Number 03260925184 Blood Gas and Electrolytes Analyzer; The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, bilirubin, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.

Recall: Z-2330-2015 · Initiated June 24, 2015

Recall

Recall Number
Z-2330-2015
Event Number
71609
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
CHL
Status
Terminated
Root Cause
Process control
Initiated
June 24, 2015
Posted
August 5, 2015
Terminated
January 12, 2016
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

S2 Fluid Pack on the Cobas b 221 system- Catalog Number 03260925184 Blood Gas and Electrolytes Analyzer; The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, bilirubin, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.

Reason

The affected S2 fluid packs may generate a transponder error when the fluid pack is loaded on to the cobas b 221 system. The error prevents the fluid from being recognized by the system and the instrument cannot start measurements until the affected S2 Fluid Pack is replaced.

Action

Roche Diagnostics sent an Urgent Medical Device Correction dated July 2, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are asked to do the following: 1. Keep in the inventory a backup S2 Fluid Packs from an unaffected lot, or at least four additional S2 Fluid Packs from the affected lots when product from unaffected lots is not available. 2. If the S2 Fluid Pack generates a Transponder error message, contact the Roche Support Network Customer Support Center at 1-800-526-2272 (provide the lot and serial number printed on the side of the affected S2 Fluid Pack). 3. After contacting Roche Support Network Customer Support Center, dispose of the affected S2 Fluid Pack per local guidelines. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272, if you have questions about the information contained in this Urgent Medical Device Correction.

Distribution

US Distribution: CT, TX, LA, ND, MI, NJ, OH, OK, PR, SC, and KY.

Quantity

19 affected devices of lot 21450781 were distributed in the US.