FDA Recall Open, Classified

Pentax Medical Video Processor; Model Number: EPK-i8020c;

Recall: Z-2329-2025 · Initiated July 16, 2025

Recall

Recall Number
Z-2329-2025
Event Number
97073
Firm
Pentax of America Inc
FEI Number
1000080301
Product Code
PEA
Status
Open, Classified
Root Cause
Device Design
Initiated
July 16, 2025
Posted
August 15, 2025
Address
3 Paragon Dr, Montvale, NJ, 07645-1782

Description

Pentax Medical Video Processor; Model Number: EPK-i8020c;

Reason

During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.

Action

On 7/16/2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Customer Instructions: PENTAX Medical will conduct the software update on the affected devices at your facility. Your local PENTAX Medical representative will contact you to schedule the required updates for your equipment. " Please download the Instructions for Use (IFU) for the EPK-i8020c Video Processor from the PENTAX Medical online IFU library at https://ifu.pentaxmedical.com. " Please complete the enclosed Field Correction Response Form upon receipt of this package, and email to PENTAX Medical at [email protected]. Contact Information: Please indicate through the attached response form that you have received and understood this information, by completing it and returning it no later than July18, 2025 at: [email protected] If you have any questions regarding this action, please feel free to contact us at: " Tel: 1-800-431-5880 (8:30 AM 5:00 PM, Monday Friday, EST) " Fax: (800)-579-5432) " Email: [email protected]

Distribution

US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.

Quantity

144 units