FDA Recall Terminated

STAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product Usage: is designed to enable complex surgery using a minimally invasive approach.

Recall: Z-2329-2020 · Initiated December 19, 2019

Recall

Recall Number
Z-2329-2020
Event Number
85556
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
KCT
Status
Terminated
Root Cause
Device Design
Initiated
December 19, 2019
Terminated
August 31, 2023
Address
1266 Kifer Rd, Bldg 101, Sunnyvale, CA, 94086-5304

Description

STAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product Usage: is designed to enable complex surgery using a minimally invasive approach.

Reason

Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential breach of sterility and risk of infection. Certain features on the trays with handling and storage may cause these holes.

Action

The firm sent a Field Safety Notice, titled Urgent Medical Device Correction Letter - ISIFA2019-10-C to customers via FedEx starting on December 19, 2019. The letter informed customers that the firm had received complaints regarding small but detectable holes in the sterilization wrap used with da Vinci Xi 8 mm Endoscope Sterilization Trays (PN 400498-02) and Handheld Camera Sterilization Trays (PN 400499-03). Customers were informed that certain features on the trays, in combination with loading and unloading of wrapped trays into sterilizers, and stacking and unstacking of wrapped trays during handling and storage could lead to these holes and a potential breach of sterility and infection in patients. Instructions were provided on steps to be taken to mitigate the potential for development of small holes, to be used in conjunction with industry practice. This included: 1. to avoid dragging a wrapped tray when loading and unloading the sterilizer; 2. to lift trays instead of sliding or dragging on and off workbenches, shelves, and other horizontal surfaces, and 3. sites using sterilizers with interior brackets were advised that they should avoid dragging trays across interior brackets. Customers were also advised to continue to inspect wrapped trays prior to use per hospital policy, and to continue to follow instructions located in the da Vinci Xi Endoscope Reprocessing Instructions and da Vinci Xi Handheld Camera Reprocessing Instructions Addendum regarding stainless steel trays. Customers were requested to inform necessary hospital personnel about this issue, complete the attached Acknowledgement Form acknowledging they had received and read the notice, ensured all personnel were fully informed of the contents, would contact Intuitive with any questions, and would retain a copy of this letter for their files. The Acknowledgement Forms were to be faxed or emailed to Intuitive Surgical as follows: ATTN: REGULATORY COMPLIANCE FIELD ACTIONS Subject line for

Distribution

Worldwide distribution - US Nationwide distribution including in the states of Arkansas, Arizona, Alabama, Alaska, California, Connecticut, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Idaho, Kentucky, Kansas, Louisiana, Missouri, Michigan, Maryland, Mississippi, Minnesota, Montana, Maine, Massachusetts, Nevada, New York, North Dakota, Nebraska, North Carolina, New Jersey, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin, Washington and West Virginia. Multiple countries including Canada.

Quantity

16 Trays