FDA Recall Terminated

Martel Printer Accessory to i-STAT Portable Clinical Analyzer, rechargeable battery pack. Printing results of clinical chemistry tests and test panels contained in the i-STAT test cartridges.

Recall: Z-2328-2009 · Initiated April 22, 2009

Recall

Recall Number
Z-2328-2009
Event Number
51872
Firm
Abbott Point of Care Inc.
FEI Number
2245578
Product Code
CGA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 22, 2009
Posted
September 30, 2009
Terminated
February 25, 2010
Address
104 Windsor Center Dr, East Windsor, NJ, 08520

Description

Martel Printer Accessory to i-STAT Portable Clinical Analyzer, rechargeable battery pack. Printing results of clinical chemistry tests and test panels contained in the i-STAT test cartridges.

Reason

It appears that the rechargeable battery pack is overheating with or without the presence of smoke.

Action

Urgent Recall Notice letters dated April 2009 were sent out the third week of April by Federal Express. The letter discussed the background, required actions, and additional information. Customers are to identify all affected Martel printer rechargeable battery packs, record the number of affected packs on the attached business reply cards, and return the card along with the affected batteries as instructed. Questions should be directed to Abbott Technical Support at 800-366-8020.

Distribution

Worldwide Distribution

Quantity

2552