FDA Recall Open, Classified

RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696

Recall: Z-2327-2025 · Initiated July 24, 2025

Recall

Recall Number
Z-2327-2025
Event Number
97318
Firm
Boston Scientific Corporation
FEI Number
2124215
Product Code
NVY
Status
Open, Classified
Root Cause
Component design/selection
Initiated
July 24, 2025
Posted
August 20, 2025
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696

Reason

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Action

On July 24, 2025, firm notified physicians via Important Medical Device Information letters. The letter provides information on the management of gradually rising daily subthreshold, low-voltage shock impedance (LVSI) pattern associated with calcification of expanded polytetrafluoroethylene (ePTFE) coated single coil (SC) and dual coil (DC) RELIANCE defibrillation leads manufactured by Boston Scientific Corporation between 2002 to 2021. Letter includes recommendations for routine follow-ups of affected leads, and considerations to take if lead replacement is planned. The firm is not recommending any changes to the scheduled follow-up interval for patients with ePTFE lead models.

Distribution

Worldwide Distribution. US nationwide. International distribution worldwide.

Quantity

247,702 US; 344,670 OUS