Infant Warmer System (IWS)

Recall: Z-2323-2017 · Initiated May 5, 2017

Recall

Recall Number
Z-2323-2017
Event Number
77287
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
FMT
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 5, 2017
Posted
May 19, 2017
Terminated
July 12, 2018
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Infant Warmer System (IWS)

Reason

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Action

GE Healthcare sent an Urgent Medical Device Correction letter dated May 5, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Product Correction : Attached to this letter, we provide instructions as part of a Service Manual Addendum on how to correct the issue. Please add this new Addendum to the Service Manual of your device(s) and train the affected users accordingly. Please acknowledge that you have received this letter and that you understand that an action needs to be taken on your part to correct this issue by filling out and returning the attached Customer Response form. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.

Distribution

Worldwide Distribution - USA (nationwide) Distribution.

Quantity

34986