FDA Recall Open, Classified

Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/169-102, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/169-101 , BD PYXIS MEDBANK MINI 1FH-2HH/155311-01, BD PYXIS MEDBANK MINI 4HH/155310-01, BD PYXIS MEDBANK MINI 1FH-1FM/155288-01, BD PYXIS MEDBANK MINI 4HM/155171-01, BD PYXIS MEDBANK MINI 3HH-1HM/155169-01, BD PYXIS MEDBANK MINI 3HH-1FG/155168-01, BD PYXIS MEDBANK MINI 2HM-1FM/155167-01, BD PYXIS MEDBANK MINI 2HH-2HM/155166-01, BD PYXIS MEDBANK MINI 2HH-1FM/155165-01, BD PYXIS MEDBANK MINI 2FM/155163-01, BD PYXIS MEDBANK MINI 1HH-3HM/155161-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/155160-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/155159-01, BD PYXIS MEDBANK MINI 2HM-1FM/139131-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/139122-01, BD PYXIS MEDBANK MINI 3HH-1FG/139091-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/138921-01, BD PYXIS MEDBANK MINI 4HM/138915-01, BD PYXIS MEDBANK MINI 1HH-3HM/138912-01, BD PYXIS MEDBANK MINI 4HH/138910-01, BD PYXIS MEDBANK MINI 3HH-1HM/138909-01, BD PYXIS MEDBANK MINI 2HH-1FH/138908-01, BD PYXIS MEDBANK MINI 2HH-2HM/138907-01, BD PYXIS MEDBANK MINI 2HH-1FM/138906-01, BD PYXIS MEDBANK MINI 2FM/138903-01, BD PYXIS MEDBANK MINI 2FH/138902-01, MEDBANK MINI MAIN/1147-00, MEDBANK MAIN/1137-00, BD PYXIS MEDBANK TWR MN 8HH-3FM/169-81, BD PYXIS MEDBANK TWR MN 8HH-3FH/169-80, BD PYXIS MEDBANK TWR MN 8HH-3FF/169-79, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/169-78, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/169-77, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/169-76, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/169-75, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/169-74, BD PYXIS MEDBANK TWR MN 7FM/169-73, BD PYXIS MEDBANK TWR MN 7FH/169-72, BD PYXIS MEDBANK TWR MN 6HH-8HM/169-71, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/169-70, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/169-69, BD PYXIS MEDBANK TWR MN 6HH-4FM/169-68, BD PYXIS MEDBANK TWR MN 6HH-4FH /169-67, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/169-66, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/169-65, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/169-64, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/169-63, BD PYXIS MEDBANK TWR MN 6FH-1FM/169-62, BD PYXIS MEDBANK TWR MN 5HH-9HM/169-61, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/169-60, BD PYXIS MEDBANK TWR MN 4HM-5FM/169-59, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/169-58, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/169-57, BD PYXIS MEDBANK TWR MN 4HH-5FM/169-56, BD PYXIS MEDBANK TWR MN 4HH-5FH /169-55, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/169-54, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/169-53, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/169-52, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/169-51, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/169-50, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/169-49, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/169-48, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/169-45, BD PYXIS MEDBANK TWR MN 4HH-10HM/169-44, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/169-43, BD PYXIS MEDBANK TWR MN 3FH-4FM/169-42, BD PYXIS MEDBANK TWR MN 2HM-6FM/169-41, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/169-40, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/169-39, BD PYXIS MEDBANK TWR MN 2HH-6FM/169-38, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/169-37, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/169-36, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/169-35, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/169-34, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/169-33, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/169-32, BD PY

Recall: Z-2321-2025 · Initiated July 10, 2025

Recall

Recall Number
Z-2321-2025
Event Number
97245
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
BRY
Status
Open, Classified
Root Cause
Use error
Initiated
July 10, 2025
Posted
August 13, 2025
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/169-102, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/169-101 , BD PYXIS MEDBANK MINI 1FH-2HH/155311-01, BD PYXIS MEDBANK MINI 4HH/155310-01, BD PYXIS MEDBANK MINI 1FH-1FM/155288-01, BD PYXIS MEDBANK MINI 4HM/155171-01, BD PYXIS MEDBANK MINI 3HH-1HM/155169-01, BD PYXIS MEDBANK MINI 3HH-1FG/155168-01, BD PYXIS MEDBANK MINI 2HM-1FM/155167-01, BD PYXIS MEDBANK MINI 2HH-2HM/155166-01, BD PYXIS MEDBANK MINI 2HH-1FM/155165-01, BD PYXIS MEDBANK MINI 2FM/155163-01, BD PYXIS MEDBANK MINI 1HH-3HM/155161-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/155160-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/155159-01, BD PYXIS MEDBANK MINI 2HM-1FM/139131-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/139122-01, BD PYXIS MEDBANK MINI 3HH-1FG/139091-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/138921-01, BD PYXIS MEDBANK MINI 4HM/138915-01, BD PYXIS MEDBANK MINI 1HH-3HM/138912-01, BD PYXIS MEDBANK MINI 4HH/138910-01, BD PYXIS MEDBANK MINI 3HH-1HM/138909-01, BD PYXIS MEDBANK MINI 2HH-1FH/138908-01, BD PYXIS MEDBANK MINI 2HH-2HM/138907-01, BD PYXIS MEDBANK MINI 2HH-1FM/138906-01, BD PYXIS MEDBANK MINI 2FM/138903-01, BD PYXIS MEDBANK MINI 2FH/138902-01, MEDBANK MINI MAIN/1147-00, MEDBANK MAIN/1137-00, BD PYXIS MEDBANK TWR MN 8HH-3FM/169-81, BD PYXIS MEDBANK TWR MN 8HH-3FH/169-80, BD PYXIS MEDBANK TWR MN 8HH-3FF/169-79, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/169-78, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/169-77, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/169-76, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/169-75, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/169-74, BD PYXIS MEDBANK TWR MN 7FM/169-73, BD PYXIS MEDBANK TWR MN 7FH/169-72, BD PYXIS MEDBANK TWR MN 6HH-8HM/169-71, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/169-70, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/169-69, BD PYXIS MEDBANK TWR MN 6HH-4FM/169-68, BD PYXIS MEDBANK TWR MN 6HH-4FH /169-67, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/169-66, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/169-65, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/169-64, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/169-63, BD PYXIS MEDBANK TWR MN 6FH-1FM/169-62, BD PYXIS MEDBANK TWR MN 5HH-9HM/169-61, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/169-60, BD PYXIS MEDBANK TWR MN 4HM-5FM/169-59, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/169-58, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/169-57, BD PYXIS MEDBANK TWR MN 4HH-5FM/169-56, BD PYXIS MEDBANK TWR MN 4HH-5FH /169-55, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/169-54, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/169-53, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/169-52, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/169-51, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/169-50, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/169-49, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/169-48, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/169-45, BD PYXIS MEDBANK TWR MN 4HH-10HM/169-44, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/169-43, BD PYXIS MEDBANK TWR MN 3FH-4FM/169-42, BD PYXIS MEDBANK TWR MN 2HM-6FM/169-41, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/169-40, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/169-39, BD PYXIS MEDBANK TWR MN 2HH-6FM/169-38, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/169-37, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/169-36, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/169-35, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/169-34, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/169-33, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/169-32, BD PY

Reason

Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.

Action

On 7/10/2025, correction notices were mailed and emailed to customers who were informed of the following: Firm strengthened its user guide content regarding manual access to medications. To access the updated content for the user guide and nursing training material, visit the MedBank Customer Portal at https://bd.com/medbankportal When selecting a medication from override/add item function, it is important for users to use extra precautions and verify correct medications are selected and pulled as undesired medications could be displayed on the override list. Firm recommends the following actions to prevent delays in access to medication: 1) Make sure staff know how to use the override/add-item workflow. 2) Have device keys available on-site in a secure location known to designated staff. 3) Train staff on how to manually access the device, see User Guide. 4) Ensure downtime policies and procedures are implemented and staff are appropriately trained Enable Over-ride permission. 5) Do not use the station as a crash cart or code box. Medication or items may not be immediately accessible for an emergent situation causing a delay in treatment which could lead to serious injury or death. 6) For pharmacies providing products to clinical users that are not physically located at the pharmacy site, ensure clinical users are aware of the risk of delay in access and provide access to keys. 7) Protocols for retrieving medications from alternative inventory locations 8) Guidance on storing time-sensitive or critical medications outside of automated dispensing devices to prevent access delays. 9) Complete and return the response form via email to [email protected] Distributors: provide a copy of this customer letter to any customers who may have received impacted devices. Firm is evaluating design enhancements to improve usability of the override/Add Item workflow. Report any device issues via phone: 866-930-9251

Distribution

US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH, FL, WA, MS, OK, TN, PA, RI, TX, SC, AR, MI, NJ, CT, MD, AZ, CA, WV, NC, UT, WY, CO, SD, ID, IL, IN, NV, NM, NE, KS, GA, WI, MO, IA, ME, LA, MN, AL, VT, NH, DE, MT, ND, AK, HI, DC.

Quantity

6076