Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

Recall: Z-2318-2017 · Initiated May 10, 2017

Recall

Recall Number
Z-2318-2017
Event Number
77251
Firm
Zimmer Biomet Spine Inc.
FEI Number
3012447612
Product Code
LYQ
Status
Terminated
Root Cause
Device Design
Initiated
May 10, 2017
Posted
May 15, 2017
Terminated
May 25, 2018
Address
10225 Westmoor Dr, Westminster, CO, 80021-2702

Description

Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

Reason

Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.

Action

Consignees will be notified on 5/15/17 via certified mail to return products.

Distribution

Nationwide, military/VA/govt and foreign consignee (No Canadian consignees)

Quantity

497 units