FDA Recall Open, Classified

iLED 7 surgical lights, Model Numbers: a) 4068110 (Ceiling Single Surgical Light); b) 4068120 (Mobile Surgical Light); c) 4068140 (Pendant Surgical Light); d) 4068210 (Ceiling Duo Surgical Lights); e) 4068310 (Ceiling Trio Surgical Lights); f) 4068410 (Ceiling Quad Surgical Lights)

Recall: Z-2315-2024 · Initiated June 4, 2024

Recall

Recall Number
Z-2315-2024
Event Number
94766
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FTD
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 4, 2024
Posted
July 5, 2024
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

iLED 7 surgical lights, Model Numbers: a) 4068110 (Ceiling Single Surgical Light); b) 4068120 (Mobile Surgical Light); c) 4068140 (Pendant Surgical Light); d) 4068210 (Ceiling Duo Surgical Lights); e) 4068310 (Ceiling Trio Surgical Lights); f) 4068410 (Ceiling Quad Surgical Lights)

Reason

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads

Action

Baxter issued an Urgent Medical Device Correction notice to its consignees on 06/04/2024 via US mail, first class. The notice explained the issue with the devices, potential hazards involved, and requested the following: "Actions to be Taken by Customers: 1. Operators may continue to use the iLED 7 surgical light systems by following the current instructions for use (IFU) cautions and warnings while considering the additional information below. . Utilize the lowest possible illumination level suitable for the procedure, especially in certain neurological or intestinal procedures on delicate, thin, dry, or abnormal tissue. . Avoid overlapping light fields that are set at high intensity (80% or higher). - If the light intensity of one light-head is set at 80% or higher the second light-head, if overlapping, should be set at 50% or lower. - If the light intensity of two or more light heads is set at 80% or higher, to minimize risk do not overlap them. . If a very high-intensity setting is temporarily required, reduce intensity as soon as the need passes. . Ensure the Adaptive Light Control (ALC) Plus sensor is activated and indicated on the wall or mobile control panel. If activation of the ALC Plus is not possible, please contact a technical service representative for further instructions. . If the ALC Plus is switched off, the light-head needs to be positioned at 100 centimeters (39.37 inches) from the surgical field. At other distances the illuminance and irradiance limits can be exceeded, which may result in thermal skin injuries. 2. Additional safety information reviewing the potential risks of using an overhead surgical light, including educational safety content is available through Pfiedler education (a division of AORN) at the URL listed below. This website requires a one-time registration. https://www.pfiedlereducation.com/diweb/gateway/f/https*3A*2F*2Fwww.pfi edlereducation.com*2Fdiweb*2Fcatalog*2Fitem*2Feid*2F1552-2024. " The notice also requested tha

Distribution

Worldwide distribution.

Quantity

6243 units