FDA Recall Terminated

Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity.

Recall: Z-2312-2012 · Initiated July 24, 2012

Recall

Recall Number
Z-2312-2012
Event Number
62982
Firm
Capintec Inc
FEI Number
2518443
Product Code
IZD
Status
Terminated
Root Cause
Software design
Initiated
July 24, 2012
Posted
September 5, 2012
Terminated
December 10, 2013
Address
620 Alpha Dr, Pittsburgh, PA, 15238

Description

Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity.

Reason

Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake protocol ?Measure Liquid and Subtract Residual." This protocol was a new feature added to Software Revision 1.22 in January 2011. The error occurs only if the user does NOT measure the residual dose immediately after me

Action

The firm, CAPINTEC, INC., sent a "SOFTWARE ERROR ALERT" letter dated July 20, 2012 to all affected consignees/customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the four measuring steps outlined in the letter; DO NOT PRINT, EXIT THE THYROID UPTAKE MODULE, OR MEASURE ANOTHER PATIENT BEFORE YOU HAVE COMPLETED THESE FOUR MEASUREMENT STEPS; and follow the attached addendum for step-by step instructions. Capintec will send corrected software to all affected customers in approximately 2-3 weeks. The customers are instructed to contact Capintec Customer Support Representative if you have any questions or concerns regarding this notification, or if you do not receive your replacement software. Capintec Customer Support can be reached by phone at 1-800-631-3826 or by email at [email protected] or by visiting our website at www.capintec.com.

Distribution

Worldwide distribution: USA (nationwide) and countries of: Argentina, Belgium, Brazil, Canada, China, Greece, Hong Kong, India, Israel, Malaysia, Philippines, Spain, South Korea, Thailand, and Turkey.

Quantity

256