FDA Recall
Open, Classified
OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004
Recall: Z-2310-2021
·
Initiated June 25, 2021
Recall
- Recall Number
- Z-2310-2021
- Event Number
- 88344
- Firm
- OPTI Medical Systems, Inc
- FEI Number
- 3004102403
- Product Code
- QJR
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 25, 2021
- Address
- 235 Hembree Park Dr, Roswell, GA, 30076-5738
Description
OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004
Reason
The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.
Action
/The firm disseminated a technical bulletin beginning on 06/25/2021 by email identifying the /problem and providing additional instructions to be implemented when using the product.
Distribution
Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados
Quantity
3552 kits