FDA Recall Open, Classified

OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Recall: Z-2310-2021 · Initiated June 25, 2021

Recall

Recall Number
Z-2310-2021
Event Number
88344
Firm
OPTI Medical Systems, Inc
FEI Number
3004102403
Product Code
QJR
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
June 25, 2021
Address
235 Hembree Park Dr, Roswell, GA, 30076-5738

Description

OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Reason

The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.

Action

/The firm disseminated a technical bulletin beginning on 06/25/2021 by email identifying the /problem and providing additional instructions to be implemented when using the product.

Distribution

Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados

Quantity

3552 kits