Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.
Recall
- Recall Number
- Z-2303-2013
- Event Number
- 66108
- Firm
- Radiometer America Inc
- FEI Number
- 1523456
- Product Code
- KHP
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 8, 2012
- Posted
- September 30, 2013
- Terminated
- December 24, 2013
- Address
- 810 Sharon Dr, Westlake, OH, 44145-1521
Description
Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
On 11/07/2012, the firm sent "Urgent Field Safety Notices to their customers. The product issue was described and recommended actions were provided. Customers may contact 1 (800) 736-0600, option 4 if questions.
Nationwide Distribution, including the states of: AK, AL, AR, AZ, BM, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.
13,632 boxes/4 membranes