FDA Recall Terminated

Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90..

Recall: Z-2301-2008 · Initiated June 30, 2008

Recall

Recall Number
Z-2301-2008
Event Number
48717
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
HWA
Status
Terminated
Root Cause
Device Design
Initiated
June 30, 2008
Posted
September 20, 2008
Terminated
November 12, 2009
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90..

Reason

The instrument may fracture during use, resulting in metal fragments being left in the patient post-surgery, which could cause implant failure.

Action

Consignees were notified via letter dated 6/30/08. The letter described the issue, provided instructions for the return of the product and asked customers to complete a reply back form. For assistance, contact Zimmer at 1-574-372-4487.

Distribution

Nationwide, Australia, China, England, France, Germany, India, Italy, Japan, Korea, Singapore and Taiwan.

Quantity

1439