FDA Recall
Terminated
Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90..
Recall: Z-2301-2008
·
Initiated June 30, 2008
Recall
- Recall Number
- Z-2301-2008
- Event Number
- 48717
- Firm
- Zimmer Inc.
- FEI Number
- 1000220733
- Product Code
- HWA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 30, 2008
- Posted
- September 20, 2008
- Terminated
- November 12, 2009
- Address
- 345 E Main St, Warsaw, IN, 46580-2746
Description
Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90..
Reason
The instrument may fracture during use, resulting in metal fragments being left in the patient post-surgery, which could cause implant failure.
Action
Consignees were notified via letter dated 6/30/08. The letter described the issue, provided instructions for the return of the product and asked customers to complete a reply back form. For assistance, contact Zimmer at 1-574-372-4487.
Distribution
Nationwide, Australia, China, England, France, Germany, India, Italy, Japan, Korea, Singapore and Taiwan.
Quantity
1439