Bone Screw 4.5MM X 52MM Bone Screw Lck-Cann, Rx only Biomet, 100 Interpace Parkway, Parsippany, NJ 070454 The screw devices are an internal fixation device intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred. Implants used in this application include non sterile and sterile bone screws.
Recall
- Recall Number
- Z-2296-2009
- Event Number
- 51730
- FEI Number
- 2242816
- Product Code
- HSB
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 2, 2009
- Posted
- September 22, 2009
- Terminated
- September 23, 2009
- Address
- EBI, L.P. 100 Interpace Pkwy, Parsippany, NJ, 07054-1149
Description
Bone Screw 4.5MM X 52MM Bone Screw Lck-Cann, Rx only Biomet, 100 Interpace Parkway, Parsippany, NJ 070454 The screw devices are an internal fixation device intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred. Implants used in this application include non sterile and sterile bone screws.
Some devices could have been made with raw material lots that may have inclusions, and could result in decreased strength of the device.
The firm sent Urgent Recall Notices to all direct accounts, hospital Risk Management and Physicians by first class mail. Letters are dated March 29, 2009, April 2, 2009 and April 16, 2009 respectively. The letter describes the problem and that that there is a possible risk of a device failure. An additional e-mail was sent on May 22, 2009 to 5 customers due to an expansion of the recall to include 4 additional items. On June 19, 2009, the firm again expanded the recall to include 2 additional products and notified customers. Questions should be directed to William Hennig, at 1.973.299.9300. ext. 1557.
Worldwide - US -NY, NJ, MI, MD, FL, GA, PR, MN, SD, OK, TX,, LA, CA, PA, OH, NV, HI, ME, NC, KS, UT, CO, TN, WV, VA, and WA. Foreign: Brazil, Colombia, Denmark, Germany, Finland, Greece, Italy, Japan, The Netherlands, Switzerland, Sweden, Portugal, Spain, United Kingdom, Saudi Arabia, and Trinidad.
4