Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Recall
- Recall Number
- Z-2295-2016
- Event Number
- 74523
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- FMT
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 22, 2016
- Terminated
- December 7, 2016
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.
Drager mailed to customers an Urgent Medical Device Recall Notification Letter dated June 22, 2016 to affected customers. The letter identified the affected product, problem information on how to operate the device and replacement of parts information. For questions regarding this letter please call 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of your Drger Babytherm 8004/8010 please contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt).
US Nationwide Distribution including Puerto Rico and the states of AK,AL,AR,CA,CO,CT,DE,FL,GA,HI,IA,IL,IN,KS,KY,LA,MA,MD,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,TN,TS,TX,UT,VA,WA,WI,WV, and WY.
1,599 units