ANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
Recall
- Recall Number
- Z-2291-2015
- Event Number
- 71212
- Firm
- The Anspach Effort, Inc.
- FEI Number
- 1045834
- Product Code
- HBE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 13, 2015
- Posted
- July 29, 2015
- Terminated
- August 25, 2016
- Address
- 4500 Riverside Dr, Palm Beach Gardens, FL, 33410-4235
Description
ANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.
Anspach sent an Urgent Notice: Medical Device Recall (Correction) letter of Anspach G1 Dissection Tools dated May 6, 2015, to all affected customers. Customers were not required to return their inventory of GI Dissection tools. Customers were instructed to complete, sign, and return the attached reply form confirming receipt of the letter by fax or email. Customers were also instructed to forward the letter as appropriate immediately. Customers with questions should contact the Complaint Handling Unit Manager at 561-494-3673 or contact their Anspach Sales Representative. For questions regarding this recall call 561-627-1080.
Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
69 tools.