FDA Recall Open, Classified

Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28

Recall: Z-2287-2025 · Initiated July 2, 2025

Recall

Recall Number
Z-2287-2025
Event Number
97185
Firm
Spark Biomedical Inc
FEI Number
3017987352
Product Code
PZR
Status
Open, Classified
Root Cause
Process control
Initiated
July 2, 2025
Posted
August 11, 2025
Address
19830 Fm 1093 Rd, Ste 401, Richmond, TX, 77407-8621

Description

Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28

Reason

Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.

Action

On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions: 1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working. 2. Current devices may continue to be used until they are replaced. 3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives. 4. Please notify all appropriate personnel at your facility that need to know about this recall.

Distribution

US Nationwide distribution.

Quantity

78 units