Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2", REF 142542 Product Usage: general surgical scissors
Recall
- Recall Number
- Z-2287-2018
- Event Number
- 80279
- Firm
- Teleflex Medical Europe Ltd Ida Business Park Athlone Ireland
- FEI Number
- 3007134520
- Product Code
- LRW
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- June 12, 2018
- Posted
- June 14, 2018
- Terminated
- April 2, 2021
Description
Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2", REF 142542 Product Usage: general surgical scissors
The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized.
The firm initiated the recall by letter on 06/12/2018. All affected consignees are instructed to take the following actions: "1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot numbers listed above. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter."
Worldwide Distribution - US Nationwide and the countries of Canada, India
573 units