FDA Recall Open, Classified

FOOT ANKLE-LF Procedure tray Catalog Number:CCFA51B

Recall: Z-2285-2024 · Initiated May 15, 2024

Recall

Recall Number
Z-2285-2024
Event Number
94765
Firm
American Contract Systems Inc
FEI Number
3008492337
Product Code
OJH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 15, 2024
Posted
July 5, 2024
Address
85 Shaffer Park Dr, Tiffin, OH, 44883-9290

Description

FOOT ANKLE-LF Procedure tray Catalog Number:CCFA51B

Reason

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

Action

American Contract Systems issued Urgent Voluntary Recall Letter (FA-2024-018) via email on 5/15/24 to Distributors. Distributors to notify end-users. Letter states reason for recall, health risk and action to take: Immediately discontinue use of the cast padding component (DISCARD ITEM 4927009046-CAST PADDING, 6" Cotton Blend Rayon) included in each kit/tray identified in ATTACHMENT 1. Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing labeling. "Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. "Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Complete the enclosed Response Form and return as soon as possible. "If credit is requested, please send an email to [email protected].

Distribution

US Nationwide distribution.

Quantity

107 units