FDA Recall Open, Classified

ARIES GBS Assay , REF 50-10021, UDI # 00840487100165

Recall: Z-2285-2019 · Initiated February 20, 2019

Recall

Recall Number
Z-2285-2019
Event Number
83470
Firm
Luminex Corporation
FEI Number
3002524000
Product Code
NJR
Status
Open, Classified
Root Cause
Process control
Initiated
February 20, 2019
Posted
August 15, 2019
Address
12212 Technology Blvd, Austin, TX, 78727-6101

Description

ARIES GBS Assay , REF 50-10021, UDI # 00840487100165

Reason

Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.

Action

On February 20, 2019, the firm sent a "Urgent: Medical Device Recall/Field Safety Notice" via email to all consignees. The customer notification informed customers of the following: 1. As an immediate action, Luminex is requesting that customers discontinue use of and discard the products listedin the recall notice immediately or by the Discontinue Use By date. Contact Luminex Global Support Services to obtain replacement material. Material that is still within its Discontinue Use By date may be used at the customers discretion. 2. Complete the Acknowledgment and Receipt form. You must complete and return this form even if you do not have any remaining product. Luminex Technical Support can assist you in completing this form. This information is essential in order to maintain recall effectiveness information required by US FDA. 3. Although no adverse events have been reported, adverse reaction or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 4. Please call the Luminex Global Support Services Team if you have any questions or concerns. Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) [email protected] CAN-0233

Distribution

US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom