FDA Recall Terminated

Dimension Enzymatic Creatinine, Siemens Material Number 10471520, UDI Number 00842768030512 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.

Recall: Z-2284-2020 · Initiated May 4, 2020

Recall

Recall Number
Z-2284-2020
Event Number
85602
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JFY
Status
Terminated
Root Cause
Device Design
Initiated
May 4, 2020
Posted
June 5, 2020
Terminated
September 20, 2021
Address
500 Gbc Dr, Newark, DE, 19702-2466

Description

Dimension Enzymatic Creatinine, Siemens Material Number 10471520, UDI Number 00842768030512 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.

Reason

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Action

On May 4, 2020, Siemens notified their customers that use the Fructosamine and Enzymatic Creatinine assays on the Atellica, ADVIA, and Dimension platforms of the potential for falsely depressed results in the presence of NAPQI. An Urgent Medical Device Correction Notice was sent to consignees in the US and an Urgent Field Safety Notice was sent to customers outside the US. The letter stated that the Limitations of the Procedure section of the Instructions for Use (IFU) will be updated with specific information related to the assay/platform. Customers were asked to complete and return the Effectiveness Check questionnaire attached to the letter within 30 days. In addition, customers were informed that field service personnel will be sent an Service Knowledge Base (SKB) describing the issue and instructing them how to deal with customer questions.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.

Quantity

8261