FDA Recall Terminated

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Recall: Z-2284-2012 · Initiated August 10, 2012

Recall

Recall Number
Z-2284-2012
Event Number
62776
Firm
Steris Corporation
FEI Number
1527821
Product Code
MED
Status
Terminated
Root Cause
Device Design
Initiated
August 10, 2012
Posted
August 28, 2012
Terminated
July 5, 2013
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Reason

STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

Action

The firm, Steris, sent a "Voluntary Field Correction: Re-labeling of QKC1692E" letter on August 10, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the steps outlined in the letter to re-label the affected QKC1692E(s) and return the completed form by September 30. 2012 to: Regulatory Affairs Specialist via [email protected] or via fax at (440) 350-7082. If you have any questions, contact Director, Regulatory Affairs at 440-392-7519.

Distribution

Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.

Quantity

64 units