FDA Recall Terminated

ABC PROBE REF 160636 5 mm Handswitching Probe, 36 cm QTY: 1 STERILE CONMED CORPORATION 525 French Rd, Utica, NY 13502 ASSEMBLED IN MEXICO ABC PROBE REF 160636 5 mm Handswitching Probe, 36 cm QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13502 ASSEMBLED IN MEXICO Usage: Electrosurgical handpiece for use as monopolar electrosurgical accessory in conjunction with CONMED electrosurgical units equipped with argon beam coagulation.

Recall: Z-2280-2010 · Initiated December 18, 2008

Recall

Recall Number
Z-2280-2010
Event Number
52030
Firm
ConMed Electrosurgery
FEI Number
3004491689
Product Code
GEI
Status
Terminated
Root Cause
Process design
Initiated
December 18, 2008
Posted
August 25, 2010
Terminated
October 22, 2010
Address
14603 E Fremont Ave, Centennial, CO, 80112-4251

Description

ABC PROBE REF 160636 5 mm Handswitching Probe, 36 cm QTY: 1 STERILE CONMED CORPORATION 525 French Rd, Utica, NY 13502 ASSEMBLED IN MEXICO ABC PROBE REF 160636 5 mm Handswitching Probe, 36 cm QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13502 ASSEMBLED IN MEXICO Usage: Electrosurgical handpiece for use as monopolar electrosurgical accessory in conjunction with CONMED electrosurgical units equipped with argon beam coagulation.

Reason

The internal electrode can protrude in front of ceramic insulator at the tip of the ABC Handpiece used with Conmed Electrosurgical units.

Action

An Urgent Device Recall notification was issued 12/18/2008, which explained the problem/risk associated with the device, the discontinuance of use and the return of the device. If you have any questions, please contact 800-552-0138, ext. 5225.

Distribution

Nationwide and International

Quantity

Dom: 184 cs (1840 units); Int'l: 47 cs (470 units)