FDA Recall Open, Classified

CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

Recall: Z-2279-2026 · Initiated April 24, 2026

Recall

Recall Number
Z-2279-2026
Event Number
98932
Firm
GE Medical Systems Information Technologies Inc
FEI Number
3012359877
Product Code
DQK
Status
Open, Classified
Root Cause
Process control
Initiated
April 24, 2026
Posted
June 1, 2026
Address
9900 W Innovation Dr, Milwaukee, WI, 53226-4856

Description

CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

Reason

Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.

Action

Firm notified consignees starting on April 24, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION." Pending corrections from GE Healthcare, customers were instructed to perform a visual inspection of the printer assembly to verify proper positioning of the ESD brush. Customers may continue to use the device if the brush is properly positioned and aligned. However, if the brush appears misaligned or displaced, stop using the device and contact a GE HealthCare Service representative.

Distribution

Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Bangladesh, Cambodia, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, South Korea, Lebanon, Libya, Lithuania, Malaysia, Myanmar, Netherlands, New Zealand, Norway, Philippines, Poland, Qatar, Romania, Saudi Arabia, Senegal, Singapore, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom.

Quantity

1835