FDA Recall
Terminated
Spectrum Medical Quantum Standard heat exchanger
Recall: Z-2277-2021
·
Initiated March 22, 2021
Recall
- Recall Number
- Z-2277-2021
- Event Number
- 88267
- Firm
- QURA S.R.L. Via Di Mezzo
- FEI Number
- 3016746283
- Product Code
- KFM
- Status
- Terminated
- Root Cause
- Material/Component Contamination
- Initiated
- March 22, 2021
- Posted
- August 12, 2021
- Terminated
- April 22, 2024
- Address
- 23 Mirandola Italy
Description
Spectrum Medical Quantum Standard heat exchanger
Reason
Product labeled as sterile may not have been adequately sterilized. In addition, these devices were imported to the US distributor, but the devices are not 510(k) cleared in the US.
Action
The firm notified their consignee of the problem by email beginning on 02/18/2021. They issued a recall noticed on 03/22/2021 by email.
Distribution
US Nationwide distribution in the state of South Carolina.
Quantity
1088 units