FDA Recall Terminated

CARESCAPE MONITOR B850 Usage: The CARESCAPE TM Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility.

Recall: Z-2275-2010 · Initiated July 16, 2010

Recall

Recall Number
Z-2275-2010
Event Number
56141
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
July 16, 2010
Posted
August 25, 2010
Terminated
September 18, 2013
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

CARESCAPE MONITOR B850 Usage: The CARESCAPE TM Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility.

Reason

GE Healthcare has become aware of four potential safety issues associated with the CARESCAPE" Monitor B850: 1. Loss of user input when using certain displays or the USB remote control. 2. Potential for a delay in treatment or missed alarm when the alarm light is used as a primary alarm source and audible alarms are turned down or off. 3. The CARESCAPE Monitor B850 will not use user entered hemo

Action

Consignees were sent on 7/16/10, a GE "Urgent Medical Device Correction" letter dated July 12, 2010. The letter was addressed to Healthcare Administrator/Risk Manger, Chief of Nursing, Director of Biomedical Engineering. They stated the Safety Issues, Affected Product Details, Product Correction, Safety Instructions and Contact Information. If you have any questions, please contact Technical Support at 1-800-558-7044.

Distribution

CA, CO, FL, GA, KY, LA, MA, MI, NE, NJ, NY, NC, OH, OR, PA, TX VA, and WI, Foreign: UNITED KINGDOM, TURKEY, SWITZERLAND, SPAIN , SLOVENIA, SINGAPORE, SAUDI ARABIA, PORTUGAL NETHERLAND, ITALY, GERMANY, FRANCE,FINLAND, DENMARK, CZECH REPUBLIC, BELGIUM, AUSTRIA, and AUSTRALIA.

Quantity

875