FDA Recall
Terminated
Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801
Recall: Z-2274-2010
·
Initiated June 7, 2010
Recall
- Recall Number
- Z-2274-2010
- Event Number
- 55933
- Firm
- Arrow International Inc
- FEI Number
- 3015859709
- Product Code
- KGZ
- Status
- Terminated
- Root Cause
- Process change control
- Initiated
- June 7, 2010
- Posted
- August 25, 2010
- Terminated
- September 29, 2012
- Address
- 2400 Bernville Road, Reading, PA, 19605
Description
Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801
Reason
Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed.
Action
June 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product.
Distribution
All affected product was distributed to one foreign consignee: Arrow Japan Ltd.
Quantity
3000 kits