TRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.
Recall
- Recall Number
- Z-2273-2012
- Event Number
- 51437
- Firm
- Alphatec Spine, Inc.
- FEI Number
- 2027467
- Product Code
- KWQ
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 29, 2012
- Posted
- August 27, 2012
- Terminated
- September 4, 2012
- Address
- 5818 El Camino Real, Carlsbad, CA, 92008-8816
Description
TRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.
On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
Alphatec Spine sent an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" letter dated May 29, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A form was attached for customers to complete and return via fax to 760-431-0289. Contact Alphatec Spine's Customer Service Department at [email protected] questions regarding this recall.)
Nationwide Distribution.