FDA Recall Terminated

Medtronic Everest 30 Disposable Inflation Device, AC3205P. Used to inflate/deflate balloon catheters as well as to monitor pressure within the balloon.

Recall: Z-2273-2010 · Initiated July 16, 2010

Recall

Recall Number
Z-2273-2010
Event Number
56276
Firm
Medtronic, Inc.
FEI Number
1220452
Product Code
MAV
Status
Terminated
Root Cause
Packaging process control
Initiated
July 16, 2010
Posted
August 25, 2010
Terminated
June 7, 2011
Address
35-37A Cherry Hill Drive, Danvers, MA, 01923-2565

Description

Medtronic Everest 30 Disposable Inflation Device, AC3205P. Used to inflate/deflate balloon catheters as well as to monitor pressure within the balloon.

Reason

Product sterility may be compromised due to breach of package seal integrity.

Action

Medtronic issued a Urgent Medical Device Recall Notice to US customer on 7/15/2010 and to foreign accounts advising of the issue and requesting that they segregate recalled product. A Medtronic Sales Rep will visit their facility to account for all affected product and initiate a credit. Foreign accounts will return product to their respective distribution centers.

Distribution

Worldwide Distribution -- USA, including the state of MI and the countries of Austria, Belgium, Czech Republic, Finland, France, Germany, Greece, Ireland, Israel, Italy, Lithuania, Pakistan, Poland, Portugal, Slovakia, Spain, Sweden, Switzerland, Ukraine, United Arab Emirates, and United Kingdom.

Quantity

1,254 units