SOMATOM Definition Edge, Model Number 10590000 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).
Recall
- Recall Number
- Z-2269-2019
- Event Number
- 83377
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 3, 2019
- Terminated
- May 10, 2024
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
SOMATOM Definition Edge, Model Number 10590000 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).
Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.
Siemens has initiated a Customer Safety Advisory Notice (CSAN) to all affected customers via CT035/19/S. Customers with valid email addresses were emailed a copy of the CSAN on 07/03/2019 via Adobe Mega Sign. The CSAN were distributed to the remaining customers without a valid email address via certified mail on 07/12/2019. Additionally, a solution to eliminate the root cause of this problem (application of software update VB20) will be distributed to all affected customers free of charge. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented.
US Nationwide distribution.