FDA Recall Terminated

Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

Recall: Z-2265-2012 · Initiated November 1, 2010

Recall

Recall Number
Z-2265-2012
Event Number
62828
Firm
Canon Inc. 30-2, Shimomaruko 3-chome, Ohta-ku Tokyo Japan
FEI Number
1000181430
Product Code
MQB
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 1, 2010
Posted
August 24, 2012
Terminated
January 7, 2014

Description

Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

Reason

Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.

Action

Canon Inc. notified sales companies including Canon USA on 9/15/2010 with Service Bulletin 10-017, and on 8/25/2011 with Service Bulletin 11-011, and on 1/20/2011 with Service Bulletin 11-001. Sales representatives were to visit customers and check and replace the screws.

Distribution

Nationwide Distribution (USA)

Quantity

468 units