EnVeo R Loading System, Model No. LS-EnVeoR-23, LS-EnVeoR-2629, LS-EnVeoR-23-C, LS-EnVeoR-2629-C. Cardiovascular valve delivery system for use with the Medtronic CoreValve Evolut R system.
Recall
- Recall Number
- Z-2264-2015
- Event Number
- 71719
- Firm
- Medtronic Cardiovascular Surgery-the Heart Valve Division
- FEI Number
- 2025587
- Product Code
- NPT
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 15, 2015
- Posted
- August 5, 2015
- Terminated
- May 20, 2016
- Address
- 1851 E Deere Ave, Santa Ana, CA, 92705-5720
Description
EnVeo R Loading System, Model No. LS-EnVeoR-23, LS-EnVeoR-2629, LS-EnVeoR-23-C, LS-EnVeoR-2629-C. Cardiovascular valve delivery system for use with the Medtronic CoreValve Evolut R system.
Medtronic is recalling the EnVeo R Loading Systems because of the presence of particulate.
An Urgent Medical Device Recall Letter dated July 2015 will be be sent to customers who purchased the EnVeo R Loading System because Medtronic has received 8 reports related to this issue out of 7347 potentially affected units. The letter informs customers of the problems identified and the actions to be taken. Customers are instructed to complete the customer confirmation certificate and to email [email protected] or fax to Medtronic at (651) 367-0612 to the attention of Customer Focused Quality.
Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kuwait, Latvia, Lithuania, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates and United Kingdom.
6,912 units total (540 units in US)