FDA Recall Terminated

syngo Imaging XS. Radiological image processing system.

Recall: Z-2264-2012 · Initiated July 18, 2012

Recall

Recall Number
Z-2264-2012
Event Number
62678
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
July 18, 2012
Posted
August 24, 2012
Terminated
November 6, 2014
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

syngo Imaging XS. Radiological image processing system.

Reason

There is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest", "Pixel Lens", "Edge Enhancement" and "Histograms".

Action

Siemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" dated July 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. For questions concerning this notice contact your local Siemens Service Rep..

Distribution

Nationwide Distribution-CA, CO, CT, FL, GA, IL, IN, IA, KS, MD, MN, MO, NE, NV, NJ, NY, NC, OH, PA, TN, TX, and WI.

Quantity

52