Two-Lumen Hemodialysis Catheterization Kit with Blue Flex Tip(R), ARROWg+ard Blue(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions for High Volume Infusions. The MAC(TM) Multi-Lumen Central Venous Access Device with ARROWg+ard Blue(R) permits venous access and catheter introduction to the central circulation. It may be inserted into the jugular, subclavian, or femoral veins. The ARROWg+ard(R) technology is intended to help provide protection against catheter-related infections. Clinical data have not been collected that demonstrate the use of the ARROWg+ard(R) antimicrobial surface in decreasing catheter-related infections for this device. It is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use.
Recall
- Recall Number
- Z-2260-2017
- Event Number
- 76937
- Firm
- Arrow International Inc
- FEI Number
- 3015859709
- Product Code
- FOZ
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- March 20, 2017
- Terminated
- December 18, 2018
- Address
- 2400 Bernville Rd, Reading, PA, 19605-9607
Description
Two-Lumen Hemodialysis Catheterization Kit with Blue Flex Tip(R), ARROWg+ard Blue(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions for High Volume Infusions. The MAC(TM) Multi-Lumen Central Venous Access Device with ARROWg+ard Blue(R) permits venous access and catheter introduction to the central circulation. It may be inserted into the jugular, subclavian, or femoral veins. The ARROWg+ard(R) technology is intended to help provide protection against catheter-related infections. Clinical data have not been collected that demonstrate the use of the ARROWg+ard(R) antimicrobial surface in decreasing catheter-related infections for this device. It is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use.
Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to [email protected]. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
Nationwide Distribution
43,394 units